Introducing the Bionime GM700S Blood Glucose Monitor, a cutting-edge device that redefines diabetes management. Created by Bionime, a leading name in healthcare technology, this monitor combines innovation and precision to empower individuals in their health journey.
The GM700S is designed for accuracy, providing reliable blood glucose readings for effective diabetes control. Its user-friendly interface ensures ease of use, making it accessible for individuals of all ages. The compact and portable design allows users to monitor their glucose levels conveniently, whether at home or on the go.
Advanced technology sets the GM700S apart, offering a seamless and pain-free testing experience. The monitor is equipped with features that enhance the user’s overall experience, from customizable settings to memory storage for tracking trends over time.
Bionime’s commitment to reliability and innovation shines through in the GM700S. This monitor goes beyond the basics, offering a comprehensive solution for proactive diabetes care. By incorporating the latest advancements in healthcare technology, Bionime empowers individuals to take control of their health with confidence.
Choose the Bionime GM700S Blood Glucose Monitor for an advanced and reliable tool in diabetes management. Experience the convenience of accurate readings, portability, and the assurance of the Bionime brand. Where precision meets innovation, GM700S stands as a beacon of proactive health management for a brighter and healthier future.
Technical Specification Of Bionime GM700S Blood Glucose Monitor
Product Name |
Blood Glucose Monitoring Syestem |
Brand |
Bionime |
Test strip |
GS700 |
Coding |
Auto-coding |
Enzyme / Measurement technology |
GDH / Electrochemical sensor |
Blood sample source |
Capillary, venous, arterial and neonatal whole blood |
Minimum sample volume |
0.75 µL |
HCT range |
Glucose concentrations ≦ 11.1mmol/L(200mg/dL), HCT range 20 – 70% |
|
Glucose concentrations > 11.1mmol/L(200mg/dL), HCT range 20 – 60% |
Measuring range |
0.6 – 33.3 mmol/L (10 – 600mg/dL) |
Meter dimensions (L x W x H) |
82.0 x 45.0 x 15.5 mm (3.22 x 1.77 x 0.61 inch) |
Display area (L x W) |
34.0 x 27.5 mm (1.09 x 1.34 inch) |
Weight (with battery) |
43.0 ± 5 g |
Power supply |
1 CR2032 battery |
Memory capacity |
500 test results with date and time |
Operating temperature |
6 – 44℃ (43 – 111°F) |
Operating relative humidity |
10 – 90% |
Meter storage conditions |
-10 – 60°C (14 – 140°F) |
Test strip storage conditions |
At 4 – 30°C (39 – 86°F), and below 90% relative humidity |
Bluetooth version (For GM700SB) |
4 |
Accuracy |
For glucose concentrations < 5.5 mmol/L (100 mg/dL), |
|
95% of measured glucose values within ± 0.83 mmol/L (15 mg/dL) of reference measurements. |
|
For glucose concentrations ≥ 5.5 mmol/L (100 mg/dL), |
|
95% of measured glucose values within ± 15% of reference measurements. |
Precision |
For glucose concentrations < 5.5 mmol/L (100 mg/dL), SD < 0.28 mmol/L (5 mg/dL) |
|
For glucose concentrations ≥ 5.5 mmol/L (100 mg/dL), CV < 5% |